American Journal of Ophthalmology
Volume 141, Issue 1 , Pages 35-43.e1, January 2006

Role of Silicon Contamination on Calcification of Hydrophilic Acrylic Intraocular Lenses

Presented in part as a paper at the ASCRS Symposium on Cataract, IOL and Refractive Surgery, Washington, DC, April 2005.

John A. Moran Eye Center, University of Utah, Salt Lake City, Utah.

Accepted 22 August 2005. published online 10 October 2005.

Purpose

To verify the presence of the element silicon on hydrophilic acrylic intraocular lenses (IOLs) explanted because of calcification.

Design

Interventional case series with clinicopathological correlation.

Methods

Twenty explanted IOLs with surface deposits (MemoryLens), and 20 with deposits mostly within their optic substance (SC60B-OUV and Aqua-Sense; 10 each) were used. After gross, microscopic, and histochemical analyses to confirm the presence of deposits, the lenses underwent scanning electron microscopy (SEM) with energy dispersive x-ray spectroscopy (EDS) for elemental composition, on the external surface of MemoryLens IOLs, and on the surface and internal substance of SC60B-OUV and Aqua-Sense IOLs. The weight percentage of the element silicon was obtained at the level of deposits, and at adjacent deposit-free areas in all lenses.

Results

Scanning electron microscopy (SEM) coupled with EDS confirmed that the composition of the deposits was calcium/phosphate in all cases. The element silicon was found in all 40 lenses, on all areas analyzed. The silicon weight percentage was higher at the level of the deposits. The presence of aluminum on five MemoryLens IOLs, and in most of the SC60B-OUV and Aqua-Sense lenses might be related to scattering from the aluminum mounting stubs used for surface analyses.

Conclusions

Silicone compounds have been implicated in the calcification of another hydrophilic acrylic design (Hydroview). They may also have a role in the calcification of other hydrophilic acrylic IOLs. Further investigation on the relationship between the presence of the element silicon and the silicone compounds found on calcified hydrophilic acrylic lenses is necessary.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Supported by the Research to Prevent Blindness Olga Keith Wiess Scholar Award (to Liliana Werner, MD, PhD).

PII: S0002-9394(05)00968-2

doi:10.1016/j.ajo.2005.08.045

American Journal of Ophthalmology
Volume 141, Issue 1 , Pages 35-43.e1, January 2006