American Journal of Ophthalmology
Volume 142, Issue 3 , Pages 375-380, September 2006

Measurements of Optic Disk Size With HRT II, Stratus OCT, and Funduscopy Are Not Interchangeable

  • Yaniv Barkana, MD

      Affiliations

    • Department of Ophthalmology, The New York Eye and Ear Infirmary, New York, New York
  • ,
  • Noga Harizman, MD

      Affiliations

    • Department of Ophthalmology, The New York Eye and Ear Infirmary, New York, New York
  • ,
  • Yariv Gerber, PhD

      Affiliations

    • Division of Cardiovascular Diseases, Mayo Clinic and Foundation, Rochester, Minnesota
  • ,
  • Jeffrey M. Liebmann, MD

      Affiliations

    • Department of Ophthalmology, Manhattan Eye, Ear, and Throat Hospital, New York, New York
    • Department of Ophthalmology, New York University Medical Center, New York, New York
  • ,
  • Robert Ritch, MD

      Affiliations

    • Department of Ophthalmology, The New York Eye and Ear Infirmary, New York, New York
    • Department of Ophthalmology, New York Medical College, Valhalla, New York
    • Corresponding Author InformationInquiries to Robert Ritch, MD, Professor and Chief, Glaucoma Service, Department of Ophthalmology, The New York Eye and Ear Infirmary, 310 East 14th Street, New York, NY 10003

Accepted 28 March 2006. published online 23 May 2006.

Purpose

To assess the interchangeability of optic disk size measurements using slit-lamp funduscopy, optical coherence tomography (OCT-3), and confocal scanning laser ophthalmoscopy (HRT-II) in clinical practice.

Design

Prospective nonrandomized clinical study.

Methods

Measurements of vertical disk diameter (VDD) were obtained with the three methods. Disk area was obtained from OCT and HRT printouts. True agreement between methods in measuring VDD was assessed using Bland-Altman graphs and 95% limits of agreement (LoA). Disks were classified as small, average, or large, and agreement between methods in this classification was assessed using κ statistics.

Results

Forty-eight patients were enrolled (mean age 53.4 ± 14.3 years). VDD (mean ± SD) was 1.58 ± 0.15, 1.70 ± 0.22, and 1.90 ± 0.24 mm with funduscopy, HRT, and OCT, respectively. Very large LoA were observed: −0.29 to 0.70 mm for OCT and HRT, −0.07 to 0.71 mm for OCT and funduscopy, and −0.29 to 0.53 mm for HRT and funduscopy. There was poor agreement (κ < 0.4) in classification of disk size as small, average, or large whether disk diameter or area was compared and using two definitions of disk size.

Conclusions

We observed a large range of differences in estimating disk size with HRT, OCT, and funduscopy. This precludes interchangeable use of these measurements in clinical practice, and does not allow simple conversion formulas to be proposed. In addition, there is poor agreement between these methods in classifying disk size as small, average, or large. At present, estimation of both absolute and relative disk size can only be defined separately for each measurement modality.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Supported in part by the Corinne Graber Research Fund of the New York Glaucoma Research Institute and the American Physicians Fellowship for Medicine in Israel.

PII: S0002-9394(06)00486-7

doi:10.1016/j.ajo.2006.03.065

American Journal of Ophthalmology
Volume 142, Issue 3 , Pages 375-380, September 2006