Fluidics and Heat Generation of Alcon Infiniti and Legacy, Bausch & Lomb Millennium, and Advanced Medical Optics Sovereign Phacoemulsification Systems

  • Michael S. Floyd, MPH
  • ,
  • Jeremy R. Valentine, BA, MD
  • ,
  • Randall J. Olson, MD

      Affiliations

    • Corresponding Author InformationInquiries to Randall J. Olson, MD, Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, 50 North Medical Drive, Salt Lake City, Utah 84132
    • Dr Olson is a consultant for Advanced Medical Optics.

Department of Ophthalmology and Visual Sciences, University of Utah Health Sciences Center, Salt Lake City, Utah.

Accepted 25 April 2006. published online 26 May 2006.

Purpose

To study heat generation, vacuum, and flow characteristics of the Alcon Infiniti and Bausch & Lomb Millennium with results compared with the Alcon Legacy and advanced medical optics (AMO) Sovereign machines previously studied.

Design

Experimental study.

Methods

Heat generation with continuous ultrasound was determined with and without a 200-g weight. Flow and vacuum were determined from 12 to 40-ml/min in 2-ml/min steps. The impact of a STAAR Cruise Control was also tested.

Results

Millennium created the most heat/20% of power (5.67 ± 0.51°C unweighted and 6.80 ± 0.80°C weighted), followed by Sovereign (4.59 ± 0.70°C unweighted and 5.65 ± 0.72°C weighted), Infiniti (2.79 ± 0.62°C unweighted and 3.96 ± 0.31°C weighted), and Legacy (1.99 ± 0.49°C unweighted and 4.27 ± 0.76°C weighted; P < .0001 for all comparisons between machines except Infiniti vs Legacy, both weighted). Flow studies revealed that Millennium Peristaltic was 17% less than indicated (P < .0001 to all other machines), and all other machines were within 3.5% of indicated. Cruise Control decreased flow by 4.1% (P < .0001 for same machine without it). Millennium Venturi had the greatest vacuum (81% more than the least Sovereign; P < .0001), and Cruise Control increased vacuum in a peristaltic machine 35% more than the Venturi system (P < .0001).

Conclusions

Percent power is not consistent in regard to heat generation, however, flow was accurate for all machines except Millennium Peristaltic. Restriction with Cruise Control elevates unoccluded vacuum to levels greater than the Venturi system tested.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Supported in part by a grant from Research to Prevent Blindness, Inc, New York, New York, to the Department of Ophthalmology and Visual Sciences, University of Utah.

PII: S0002-9394(06)00539-3

doi:10.1016/j.ajo.2006.04.042

American Journal of Ophthalmology
Volume 142, Issue 3 , Pages 387-392.e2, September 2006