Volume 142, Issue 4 , Pages 572.e1-572.e6, October 2006
Optimization of Anterior Eye Fluorescein Viewing
Purpose
To optimize anterior eye fluorescein viewing and image capture.
Design
Prospective experimental investigation.
Methods
The spectral radiance of ten different models of slit-lamp blue luminance and the spectral transmission of three barrier filters were measured. Optimal clinical instillation of fluorescein was evaluated by a comparison of four different instillation methods of fluorescein into 10 subjects. Two methods used a floret, and two used minims of different concentration. The resulting fluorescence was evaluated for quenching effects and efficiency over time.
Results
Spectral radiance of the blue illumination typically had an average peak at 460 nm. Comparison between three slit-lamps of the same model showed a similar spectral radiance distribution. Of the slit-lamps examined, 8.3% to 50.6% of the illumination output was optimized for >80% fluorescein excitation, and 1.2% to 23.5% of the illumination overlapped with that emitted by the fluorophore. The barrier filters had an average cut-off at 510 to 520 nm. Quenching was observed for all methods of fluorescein instillation. The moistened floret and the 1% minim reached a useful level of fluorescence in on average ∼20s (∼2.5× faster than the saturated floret and 2% minim) and this lasted for ∼160 seconds.
Conclusions
Most slit-lamps’ blue light and yellow barrier filters are not optimal for fluorescein viewing and capture. Instillation of fluorescein using a moistened floret or 1% minim seems most clinically appropriate as lower quantities and concentrations of fluorescein improve the efficiency of clinical examination.
To access this article, please choose from the options below
PII: S0002-9394(06)00573-3
doi:10.1016/j.ajo.2006.04.062
© 2006 Elsevier Inc. All rights reserved.
Volume 142, Issue 4 , Pages 572.e1-572.e6, October 2006