American Journal of Ophthalmology
Volume 142, Issue 4 , Pages 583-590 , October 2006

Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Glaucoma Patients and Healthy Subjects

  • Lorenz Barleon, MD

      Affiliations

    • Department of Ophthalmology, University of Mainz, Mainz, Germany
  • ,
  • Esther M. Hoffmann, MD

      Affiliations

    • Department of Ophthalmology, University of Mainz, Mainz, Germany
  • ,
  • Manfred Berres, PhD

      Affiliations

    • Department of Medical Biometrics, University of Mainz, Mainz, Germany
    • RheinAhrCampus Remagen, Department of Mathematics and Technology, Koblenz, Germany
  • ,
  • Norbert Pfeiffer, MD

      Affiliations

    • Department of Ophthalmology, University of Mainz, Mainz, Germany
  • ,
  • Franz H. Grus, MD, PhD

      Affiliations

    • Department of Ophthalmology, University of Mainz, Mainz, Germany
    • Corresponding Author InformationInquiries to Franz H. Grus, MD, PhD, Department of Ophthalmology, University of Mainz, 55101 Mainz, Germany

,Accepted 6 May 2006.

References 

  1. Brandt JD . Corneal thickness in glaucoma screening, diagnosis, and management . Curr Opin Ophthalmol . 2004;15:85–89
  2. Dielemans I , Vingerling JR , Hofman A , et al.   Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studies . Graefes Arch Clin Exp Ophthalmol . 1994;232:141–144
  3. Goldmann H , Schmidt T . Über Applanationsttonometrie . Ophthalmologica . 1957;134:221–242
  4. Wessels IF , Oh Y . Tonometer utilization, accuracy, and calibration under field conditions . Arch Ophthalmol . 1990;108:1709–1712
  5. Whitacre MM , Stein R . Sources of error with use of Goldmann-type tonometers . Surv Ophthalmol . 1993;38:1–30
  6. Wolfs RC , Klaver CC , Vingerling JR , et al.   Distribution of central corneal thickness and its association with intraocular pressure: The Rotterdam Study . Am J Ophthalmol . 1997;123:767–772
  7. Ehlers N , Bramsen T , Sperling S . Applanation tonometry and central corneal thickness . Acta Ophthalmol (Copenh) . 1975;53:34–43
  8. Whitacre MM , Stein RA , Hassanein K . The effect of corneal thickness on applanation tonometry . Am J Ophthalmol . 1993;115:592–596
  9. Hoffmann EM , Grus FH , Pfeiffer N . Intraocular pressure and ocular pulse amplitude using dynamic contour tonometry and contact lens tonometry . BMC Ophthalmol . 2004;4:4
  10. Kniestedt C , Nee M , Stamper RL . Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states . Graefes Arch Clin Exp Ophthalmol . 2005;243:359–366
  11. Altman DG , Bland JM . Comparison of methods of measuring blood pressure . J Epidemiol Commun Health . 1986;40:274–277
  12. Bron AM , Creuzot-Garcher C , Goudeau-Boutillon S , et al.   Falsely elevated intraocular pressure due to increased central corneal thickness . Graefes Arch Clin Exp Ophthalmol . 1999;237:220–224
  13. Shah S , Chatterjee A , Mathai M , et al.   Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic . Ophthalmology . 1999;106:2154–2160
  14. Stodtmeister R . Applanation tonometry and correction according to corneal thickness . Acta Ophthalmol Scand . 1998;76:319–324
  15. Duba I , Wirthlin AC . Dynamic contour tonometry for post-LASIK intraocular pressure measurements . Klin Monatsbl Augenheilkd . 2004;221:347–350
  16. Kaufmann C , Bachmann LM , Thiel MA . Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis . Invest Ophthalmol Vis Sci . 2003;44:3790–3794
  17. Siganos DS , Papastergiou GI , Moedas C . Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK . J Cataract Refract Surg . 2004;30:746–751
  18. Argus WA . Ocular hypertension and central corneal thickness . Ophthalmology . 1995;102:1810–1812
  19. Brandt JD , Beiser JA , Kass MA , et al.   Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS) . Ophthalmology . 2001;108:1779–1788
  20. Shimmyo M , Ross AJ , Moy A , et al.   Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African-Americans . Am J Ophthalmol . 2003;136:603–613
  21. Kanngiesser HE , Kniestedt C , Robert YC . Dynamic contour tonometry: presentation of a new tonometer . J Glaucoma . 2005;14:344–350
  22. Kaufmann C , Bachmann LM , Thiel MA . Comparison of dynamic contour tonometry with Goldmann applanation tonometry . Invest Ophthalmol Vis Sci . 2004;45:3118–3121
  23. Kniestedt C , Nee M , Stamper RL . Dynamic contour tonometry: a comparative study on human cadaver eyes . Arch Ophthalmol . 2004;122:1287–1293
  24. Ku JY , Danesh-Meyer HV , Craig JP , et al.   Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry . Eye . 2005; Forthcoming
  25. Doyle A , Lachkar Y . Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness . J Glaucoma . 2005;14:288–292
  26. Kamppeter BA , Jonas JB . Dynamic contour tonometry for intraocular pressure measurement . Am J Ophthalmol . 2005;140:318–320
  27. Kniestedt C , Lin S , Choe J , et al.   Clinical comparison of contour and applanation tonometry and their relationship to pachymetry . Arch Ophthalmol . 2005;123:1532–1557
  28. Kotecha A , White ET , Shewry JM , et al.   The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry . Br J Ophthalmol . 2005;89:1572–1575
  29. Swiss-Microtechnology-AG. PASCAL. Dynamic contour tonometer. User manual, Version 2.0, 2004:1–61.
  30. Kanski JJ . In: Klinische Ophthalmologie, Lehrbuch und Atlas, 5 . Auflage ed.. München: Urban und Fischer; 2004;p. 1–732
  31. Bland JM , Altman DG . Statistical methods for assessing agreement between two methods of clinical measurement . Lancet . 1986;1:307–310
  32. Ludbrook J . Statistical techniques for comparing measurers and methods of measurement: a critical review . Clin Exp Pharmacol Physiol . 2002;29:527–536
  33. Pache M , Wilmsmeyer S , Lautebach S , et al.   Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study . Graefes Arch Clin Exp Ophthalmol . 2005;243:763–767
  34. Gange SJ , Linton KL , Scott AJ , et al.   A comparison of methods for correlated ordinal measures with ophthalmic applications . Stat Med . 1995;14:1961–1974
  35. Hand D , Taylor C . In: Multivariate analysis of variance and repeated measures: a practical approach for behavioral scientists . Boca Raton, Florida: Chapman and Hall; 1987;p. 1–304

 Drs Grus and Hoffmann have received research support (instrument) from Swiss Microtechnology AG, Ziemer Ophthalmic Systems Group Company, Switzerland.

PII: S0002-9394(06)00610-6

doi: 10.1016/j.ajo.2006.05.030

American Journal of Ophthalmology
Volume 142, Issue 4 , Pages 583-590 , October 2006