Volume 142, Issue 5 , Pages 750-756.e3, November 2006
Wavefront Analysis and Contrast Sensitivity of Aspheric and Spherical Intraocular Lenses: A Randomized Prospective Study
Purpose
To compare visual performance, total and high order wavefront aberrations (coma, spherical aberration, and other terms), and contrast sensitivity in 120 eyes implanted with one monofocal aspheric intraocular lens (IOL) and two spherical IOLs.
Design
Randomized prospective study.
Methods
Sixty patients were randomized to receive three IOL types: Alcon AcrySofIQ (40 eyes), AcrySofNatural (40 eyes), and advanced medical optic (AMO)Sensar (40 eyes). Complete ophthalmologic examination including uncorrected visual acuity (UCVA), best-spectacle corrected visual acuity (BSCVA), corneal topography, and wavefront analysis were performed preoperatively, 30 days, and 90 days postoperatively. Pelli-Robson chart test and functional acuity contrast testing (FACT-Optec6500) were performed approximately 50 days after surgery. Statistical analyses were performed using analysis χ2, analysis of variance (ANOVA), and multiple comparisons Tukey test.
Results
After 90 days, all eyes had postoperative BSCVA ≥20/32. The AcrySofIQ IOL showed statistically significant less induction of spherical aberration (P < .001) when compared with the AMOSensar and the AcrySofNatural IOLs. The AMOSensar presented significantly less spherical aberration then the AcrySofNatural (P < .05). The AcrySofIQ also had lower values of total and high-order aberration (HOA) (P < .05) when compared with the AMOSensar and the AcrySofNatural. The mean values of trefoil 9, coma, and HOA root mean square (RMS) decreased between one and three months (P < .001, P < .001, P = .023, P < .001, respectively) in all groups. Mean Pelli-Robson contrast sensitivity values in photopic condition were similar between the groups. The AcrySofIQ showed better results in 3cpd spatial frequency in mesopic condition using FACT-Optec 6500 (P = .008), although there were no statistical differences in photopic and mesopic with glare conditions.
To access this article, please choose from the options below
Supplemental Material available at AJO.com.
PII: S0002-9394(06)00726-4
doi:10.1016/j.ajo.2006.06.031
© 2006 Elsevier Inc. All rights reserved.
Volume 142, Issue 5 , Pages 750-756.e3, November 2006