American Journal of Ophthalmology
Volume 146, Issue 1 , Pages 42-45, July 2008

Real-Time Optical Coherence Tomography-Guided Femtosecond Laser Sub-Bowman Keratomileusis on Human Donor Eyes

  • Omid Kermani

      Affiliations

    • Augenklinik am Neumarkt, Köln, Germany
    • Corresponding Author InformationInquiries to Omid Kermani, Augenklinik am Neumarkt, Schildergasse 107-109 D-50667 Köln, Germany
  • ,
  • Will Fabian

      Affiliations

    • Rowiak GmbH, Hannover, Germany
  • ,
  • Holger Lubatschowski

      Affiliations

    • Rowiak GmbH, Hannover, Germany
    • Laser Zentrum Hannover e.V., Hannover, Germany.

Accepted 3 March 2008. published online 09 April 2008.

Purpose

To determine the usefulness of optical coherence tomography (OCT) as a tool in guiding the femtosecond (fs) laser in the creation of a sub-Bowman keratomileusis (SBK) flap in human eyes.

Design

A nonrandomized case series.

Methods

In a private research laboratory setting, we performed an in vitro investigation on human autopsy eyes. Five human cadaver eyes, unsuitable for transplantation, underwent flap creation with a fs laser. The laser procedure was controlled in real-time with an OCT system (Thorlabs HL AG, Luebeck, Germany) to ensure that the cut was placed just underneath the Bowman layer. The fs laser worked at a repetition rate of 10 MHz with a single-pulse duration of < 400 fs (pulse energy in the nanoJoule range). As a control, all eyes underwent histologic dissection (toluidine blue) and were examined using light microscopy (LM).

Results

Video monitoring of the flap creation supported the feasibility of real-time OCT monitoring of the fs laser flap creation process. A clear distinction of the corneal epithelium was possible in all eyes. The Bowman membrane was not identified in all donor eyes at the given resolution of the OCT used in this study. Still, LM examination confirmed that the real-time monitoring assured a positioning of the cutting plane at minimum distance underneath the Bowman layer.

Conclusion

This small laboratory test offers evidence that real-time OCT monitoring of creation of a SBK flap using a fs laser is possible, thus ensuring that the flap is created at the proper depth.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0002-9394(08)00168-2

doi:10.1016/j.ajo.2008.03.003

American Journal of Ophthalmology
Volume 146, Issue 1 , Pages 42-45, July 2008