American Journal of Ophthalmology
Volume 149, Issue 1 , Pages 18-31.e2, January 2010

Spectral-Domain Optical Coherence Tomography: A Comparison of Modern High-Resolution Retinal Imaging Systems

  • Daniel F. Kiernan

      Affiliations

    • Department of Ophthalmology & Visual Sciences, Illinois Eye and Ear Infirmary, University of Illinois at Chicago, Chicago, Illinois
  • ,
  • William F. Mieler

      Affiliations

    • Department of Ophthalmology & Visual Sciences, Illinois Eye and Ear Infirmary, University of Illinois at Chicago, Chicago, Illinois
  • ,
  • Seenu M. Hariprasad

      Affiliations

    • Section of Ophthalmology and Visual Sciences, Department of Surgery, University of Chicago, Chicago, Illinois
    • Corresponding Author InformationInquires to Seenu M. Hariprasad, Vitreoretinal Service, Section of Ophthalmology, Department of Surgery, University of Chicago, 5841 South Maryland Avenue, MC 2114, Chicago, IL 60637

Accepted 28 August 2009.

Purpose

To provide a review of commercially available spectral-domain optical coherence tomography (SD OCT) systems in clinical ophthalmology.

Design

Perspective.

Methods

Review of current manufacturer information, selected articles from the literature, and the authors' clinical experience.

Results

Because the premise of SD OCT technology is the nonproprietary mathematical formula of Fourier transformation, 9 different SD OCT systems currently are or soon will be commercially available. Also demonstrated are Cirrus en face C-scan visualization of photoreceptor attenuation resulting from acute zonal occult outer retinopathy and Spectral OCT/scanning laser ophthalmoscopy microperimetric analysis of a macular caldera lesion resulting from North Carolina macular dystrophy.

Conclusions

Advances in high-resolution imaging of the anterior and posterior segment have revealed new in vivo details of anatomic, physiologic, and pathologic data for the practice of ophthalmology. Compared with time-domain OCT, SD OCT tends to derive increased retinal thickness and decreased nerve fiber layer thickness measurements. This is likely because of increased depth of resolution and greater volume of data acquired with each scan. Interdevice comparison is not practical because of differences in individual segment boundary algorithms. Improvements in photoreceptor inner segment–outer segment layer visualization should aid our understanding of visual dysfunction in a variety of retinal pathologic features. As the technology develops, SD OCT will continue to provide new insights about ocular structure and disease.

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 31.50 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

PII: S0002-9394(09)00663-1

doi:10.1016/j.ajo.2009.08.037

American Journal of Ophthalmology
Volume 149, Issue 1 , Pages 18-31.e2, January 2010