Effect of Accommodation on Vaulting and Movement of Posterior Chamber Phakic Lenses in Eyes With Implantable Collamer Lenses

      Purpose

      To investigate and compare vaulting and movement changes during accommodation in eyes with the V4c and V4 implantable collamer lenses (ICL).

      Design

      Comparative, observational case series.

      Methods

      The medical records of 35 eyes (18 patients) with the V4 ICL and 51 eyes (26 patients) with the V4c ICL were retrospectively examined and included in analyses. Anterior chamber depth (ACD), posterior corneal surface–to–ICL distance (endo-ICL distance), pupil size, and postoperative vaulting were evaluated using the Visante anterior chamber optical coherence tomography system. Images were taken during the nonaccommodative and accommodative states 3 months after ICL implantation. Refractive error, keratometry values, axial length, intraocular pressure, and central corneal thickness were evaluated at 3 months postoperatively.

      Results

      ICL vaulting did not significantly change during accommodation in eyes with either the V4 or V4c ICL (P = .532 for V4 ICL and P = .415 for V4c ICL). However, significant reductions in ACD, endo-ICL distance, and pupil size were observed during accommodation in both groups. In eyes with a V4 ICL, the change in [Δ] ACD was 0.2 ± 0.1 mm (P < .001), Δendo-ICL distance was 0.2 ± 0.1 mm (P < .001), and Δpupil size was 0.5 ± 0.9 mm (P = .021). For eyes with the V4c ICL, ΔACD was 0.2 ± 0.2 mm (P < .001), Δendo-ICL distance was 0.2 ± 0.2 mm (P < .001), and Δpupil size was 0.8 ± 1.2 mm (P < .001). The mean reductions of each parameter were not statistically different between eyes with the V4 ICL and the V4c ICL.

      Conclusions

      Contrary to the light stimulation response, accommodation does not significantly affect ICL vaulting differently in eyes with either the V4 or V4C ICLs.
      To read this article in full you will need to make a payment

      References

        • Igarashi A.
        • Shimizu K.
        • Kamiya K.
        Eight-year follow-up of posterior chamber phakic intraocular lens implantation for moderate to high myopia.
        Am J Ophthalmol. 2014; 157: 532-539.e531
        • Sanders D.R.
        • Doney K.
        • Poco M.
        United States Food and Drug Administration clinical trial of the Implantable Collamer Lens (ICL) for moderate to high myopia: three-year follow-up.
        Ophthalmology. 2004; 111: 1683-1692
        • Sanders D.R.
        • Vukich J.A.
        • Doney K.
        • Gaston M.
        U.S. Food and Drug Administration clinical trial of the Implantable Contact Lens for moderate to high myopia.
        Ophthalmology. 2003; 110: 255-266
        • Sanders D.R.
        • Brown D.C.
        • Martin R.G.
        • Shepherd J.
        • Deitz M.R.
        • DeLuca M.
        Implantable contact lens for moderate to high myopia: phase 1 FDA clinical study with 6 month follow-up.
        J Cataract Refract Surg. 1998; 24: 607-611
        • Uusitalo R.J.
        • Aine E.
        • Sen N.H.
        • Laatikainen L.
        Implantable contact lens for high myopia.
        J Cataract Refract Surg. 2002; 28: 29-36
        • Schmidinger G.
        • Lackner B.
        • Pieh S.
        • Skorpik C.
        Long-term changes in posterior chamber phakic intraocular collamer lens vaulting in myopic patients.
        Ophthalmology. 2010; 117: 1506-1511
        • Pineda-Fernandez A.
        • Jaramillo J.
        • Vargas J.
        • Jaramillo M.
        • Galindez A.
        Phakic posterior chamber intraocular lens for high myopia.
        J Cataract Refract Surg. 2004; 30: 2277-2283
        • Gonvers M.
        • Bornet C.
        • Othenin-Girard P.
        Implantable contact lens for moderate to high myopia: relationship of vaulting to cataract formation.
        J Cataract Refract Surg. 2003; 29: 918-924
        • Alfonso J.F.
        • Lisa C.
        • Abdelhamid A.
        • Fernandes P.
        • Jorge J.
        • Montes-Mico R.
        Three-year follow-up of subjective vault following myopic implantable collamer lens implantation.
        Graefes Arch Clin Exp Ophthalmol. 2010; 248: 1827-1835
        • Brandt J.D.
        • Mockovak M.E.
        • Chayet A.
        Pigmentary dispersion syndrome induced by a posterior chamber phakic refractive lens.
        Am J Ophthalmol. 2001; 131: 260-263
        • Chen L.J.
        • Chang Y.J.
        • Kuo J.C.
        • Rajagopal R.
        • Azar D.T.
        Metaanalysis of cataract development after phakic intraocular lens surgery.
        J Cataract Refract Surg. 2008; 34: 1181-1200
        • Du C.
        • Wang J.
        • Wang X.
        • Dong Y.
        • Gu Y.
        • Shen Y.
        Ultrasound biomicroscopy of anterior segment accommodative changes with posterior chamber phakic intraocular lens in high myopia.
        Ophthalmology. 2012; 119: 99-105
        • Huseynova T.
        • Ozaki S.
        • Ishizuka T.
        • Mita M.
        • Tomita M.
        Comparative study of 2 types of implantable collamer lenses, 1 with and 1 without a central artificial hole.
        Am J Ophthalmol. 2014; 157: 1136-1143
        • Alfonso J.F.
        • Lisa C.
        • Fernandez-Vega Cueto L.
        • Belda-Salmeron L.
        • Madrid-Costa D.
        • Montes-Mico R.
        Clinical outcomes after implantation of a posterior chamber collagen copolymer phakic intraocular lens with a central hole for myopic correction.
        J Cataract Refract Surg. 2013; 39: 915-921
        • Shimizu K.
        • Kamiya K.
        • Igarashi A.
        • Shiratani T.
        Intraindividual comparison of visual performance after posterior chamber phakic intraocular lens with and without a central hole implantation for moderate to high myopia.
        Am J Ophthalmol. 2012; 154: 486-494.e481
        • Higueras-Esteban A.
        • Ortiz-Gomariz A.
        • Gutierrez-Ortega R.
        • et al.
        Intraocular pressure after implantation of the Visian Implantable Collamer Lens With CentraFLOW without iridotomy.
        Am J Ophthalmol. 2013; 156: 800-805
        • Lee H.
        • Kang S.Y.
        • Seo K.Y.
        • et al.
        Dynamic vaulting changes in V4c versus V4 posterior chamber phakic lenses under differing lighting conditions.
        Am J Ophthalmol. 2014; 158: 1199-1204.e1191
        • Lindland A.
        • Heger H.
        • Kugelberg M.
        • Zetterstrom C.
        Vaulting of myopic and toric Implantable Collamer Lenses during accommodation measured with Visante optical coherence tomography.
        Ophthalmology. 2010; 117: 1245-1250
        • Petternel V.
        • Koppl C.M.
        • Dejaco-Ruhswurm I.
        • Findl O.
        • Skorpik C.
        • Drexler W.
        Effect of accommodation and pupil size on the movement of a posterior chamber lens in the phakic eye.
        Ophthalmology. 2004; 111: 325-331
        • Schachar R.A.
        • Koivula A.
        The stress on the anterior lens surface during human in vivo accommodation.
        Br J Ophthalmol. 2008; 92: 348-350
        • Koretz J.F.
        • Cook C.A.
        • Kaufman P.L.
        Accommodation and presbyopia in the human eye. Changes in the anterior segment and crystalline lens with focus.
        Invest Ophthalmol Vis Sci. 1997; 38: 569-578
        • Baikoff G.
        • Lutun E.
        • Ferraz C.
        • Wei J.
        Static and dynamic analysis of the anterior segment with optical coherence tomography.
        J Cataract Refract Surg. 2004; 30: 1843-1850
        • Dubbelman M.
        • Van der Heijde G.L.
        • Weeber H.A.
        Change in shape of the aging human crystalline lens with accommodation.
        Vision Res. 2005; 45: 117-132
        • Richdale K.
        • Bullimore M.A.
        • Zadnik K.
        Lens thickness with age and accommodation by optical coherence tomography.
        Ophthalmic Physiol Opt. 2008; 28: 441-447
        • Tsubota K.
        • Nakamori K.
        Dry eyes and video display terminals.
        N Engl J Med. 1993; 328: 584
        • Wu S.
        • He L.
        • Li J.
        • Wang J.
        • Wang S.
        Visual display terminal use increases the prevalence and risk of work-related musculoskeletal disorders among Chinese office workers: a cross-sectional study.
        J Occup Health. 2012; 54: 34-43
        • Glasser A.
        • Kaufman P.L.
        The mechanism of accommodation in primates.
        Ophthalmology. 1999; 106: 863-872
        • Gambra E.
        • Ortiz S.
        • Perez-Merino P.
        • Gora M.
        • Wojtkowski M.
        • Marcos S.
        Static and dynamic crystalline lens accommodation evaluated using quantitative 3-D OCT.
        Biomed Opt Express. 2013; 4: 1595-1609

      Biography

      Hun Lee, MD, is currently Assistant professor of Department of Ophthalmology in Catholic Kwandong University College of Medicine, International St. Mary's Hospital, Incheon, South Korea. He graduated from Yonsei University College of Medicine in 2006, where he completed his ophthalmology residency in 2011 and corneal fellowship in 2015. His research interests are visual quality and refractive error, including cataract and refractive surgery.